Natural gas - Determination of composition and associated uncertainty by gas chromatography - Part 5: Isothermal method for nitrogen, carbon dioxide, C1 to C5 hydrocarbons and C6+ hydrocarbons
Committee: Petroleum Products, Fuels, and Lubricants
Sector: Chemicals and Petroleum
Available Standards
Original Standard
إنجليزي (English) 34 Page 2025
International Classification for Standards (ICS)
Code Classification
75.060 Natural gas
Adopted Standards
Standards / Amendment / Corrigendum
1 ISO 6974-5:2014
Replacement Sequence
Scope
This part of this International Standard describes a gas chromatographic method for the quantitative determination of the content of nitrogen, carbon dioxide and C1 to C5 hydrocarbons individually and a composite C6+ measurement, which represents all hydrocarbons of carbon number 6 and above in natural gas samples. It is applicable to the analysis of gases containing constituents within the working ranges given in Table 1.
Table 1 — Component working ranges
Component

Mole fraction

%
Min.Max.
NitrogenN20,122
Carbon dioxideCO20,0515
MethaneCH434100
EthaneC2H60,123
PropaneC3H80,0510
iso-Butanei-C4H100,012,0
n-Butanen-C4H100,012,0
neo-Pentaneneo-C5H120,0050,35
iso-Pentanei-C5H120,0050,35
n-Pentanen-C5H120,0050,35
Hexanes +C6+0,0050,35
NOTE 1 The working ranges in Table 1 are those for which the method has been shown to be satisfactory, and are offered for guidance. However, there is no reason why wider ranges should not be used, provided that the successful measurement of such wider ranges has been demonstrated.
NOTE 2 Hydrocarbons above n-pentane are expressed as the “pseudo-component” C6+ which is measured as one composite peak and calibrated as such. The properties of C6+ are calculated from an extended analysis of the individual C6 and higher hydrocarbons.
NOTE 3 Oxygen is not a normal constituent of natural gas and would not be expected to be present in gas sampled to an online instrument. If any oxygen is present as a result of air contamination, it will be measured with the nitrogen. The resulting measured (nitrogen + oxygen) value will be in error to a small extent because of the slight difference between the detector responses of oxygen and nitrogen.
NOTE 4 The helium and argon contents are assumed to be sufficiently small and unvarying that they need not be analysed for.
NOTE 5 The gas sample shall not contain any hydrocarbon condensate and/or water.